[2025] VCC 1909
General information, not legal advice.
County Court of Victoria
2025-12-16
Original judgment source is linked above.
Sentence: Fine - $230,000, with a stay of 2 months granted : Section 6AAA Sentencing Act 1991 declaration: but for the Plea of guilty a fine of $330,000 would have been imposed.
(a) assessed for safe functionality prior to first use with the assessment for that chair recorded; and
(b) thereafter listed for preventative maintenance on a regular basis with each subsequent assessment recorded; and
(c) tagged or stickered with the frequency of assessment and the date of the last assessment to enable any employees to identify whether a further assessment was required.
iv. Benalla Health failed to provide and maintain the above system.
v. Persons other than Benalla Health's employees were exposed to risk including but not limited to Lorna Exton.
96-year-old husband Bernie who was also a resident at the aged care facility, with her son, Greg and his wife, Jenny. When Greg and Jenny Exton arrived at the facility, Lorna and Bernie were positioned in their chairs under the shade of a tree on a section of fake grass next to the fence. A staff member, Maria Cook, had positioned Mrs Exton for the fence visit in a reclining chair. The chair was from the acute ward in the Benalla Hospital. The chair had been at the facility for at least a couple of weeks and had been obtained because Mrs Exton had recently declined in her abilities. 4. At the conclusion of the hour-long visit, Greg Exton indicated to the staff that the visit was concluded. A staff member came over to both Mr and Mrs Exton to move them back inside the facility. The staff member stood behind Mrs Exton's chair and attempted to pull it back onto the paved walkway, however the chair did not move. The staff member inspected the front of the chair before returning to the back and pulling on the chair again. Another staff member pulled on the back of the chair. It completely separated from the base of the chair, causing Mrs Exton to fall backwards into the broken back section of the chair, which at this stage was on the ground. 5. There were two metal rods protruding vertically from the chair and they hit Mrs Exton's back. This made her visibly and audibly distressed, resulting in more staff arriving to assist her. A manual lifting device was used to lift her from the ground, and she was taken back inside the facility for assessment. An x-ray revealed a displaced fracture of the axillary portion of her right ninth rib and compression fractures of L1 and L3. Those later injuries are perhaps associated with osteoporosis and not the fall. 6. Over the next 48 hours, Mrs Exton was in significant pain and her mobility was restricted. She was prescribed opioids to reduce the pain and she was placed in palliative care on an end-of-life pathway on about 23 October and passed away on 27 October. 7. The facility notified WorkSafe of the incident. WorkSafe inspectors determined that the chair involved was a Montego air cool high care mobile battered powered chair purchased on 31 October 2014. The chair was decommissioned and disposed of after the incident. It appeared that the tension screws had loosened in it. Mexop is an online system used to schedule maintenance requests and inspections of the aged care facility. No maintenance, maintenance request or service history were located in that system relating to the chair. 8. The maintenance of equipment was the responsibility of Benalla Health, its engineering department. Equipment coming into the facility should have been put on a maintenance schedule, but there was no tagging system to let staff know when an item of equipment had been last assessed. Staff in the acute ward of the Benalla Hospital undertake a two-day orientation process to familiarise themselves with policies, processes and undertake buddy shifts at the commencement of their employment. They know to check the tension screws prior to using the chair. However, the aged care facility staff had borrowed the chair from the acute ward and had not undergone that training process. 9. On inspection by staff, it appeared that the tension screws and rod had worn on the chair. Inspectors were also supplied with email correspondence between Benalla Health and Fresenius Medical Care that noted that the chair was originally supplied with a strap with a clip lock around the back and seat frames to stop the back coming completely free. No strap-like device was on the chair on the day of the incident. If the screws were not sufficiently tight, there was nothing to prevent the backrest from being pulled free if force was applied to it, especially if the screws had worn and they were not tight enough. 10. It was reasonably practicable for Benalla Health to implement measures to eliminate or reduce the risk including the implementation of a system for the assessment of ongoing inspection and maintenance or all chairs used to transport residents. It was reasonably practical to ensure that each chair used to transport residents was:
a) assessed for safety functionality prior to use and the assessment of that chair being recorded; > > b) thereafter, listed for preventative maintenance on a regular basis and each subsequent assessment recorded; and > > c) a tag or stickering process with frequency of assessment, date of last assessment, listed to enable the employees to identify whether further assessment was requested.
[5] DPP v Frewstal Pty Ltd [2015] VSCA 266 - 47 VR 660
DPP
Benalla Health [2025] VCC 1909