TABLE OF CONTENTS
The claims in the proceedings
Decision
Ventilators
The Shangrila Ventilator and the VG70 Ventilator
Events leading up to the Agreements
HealthShare's decision to purchase 100 VG70 Ventilators from RAPP
Therapeutic Goods Administration approvals
Consideration of "lower performance options"
Further discussions between HealthShare and MDT
HealthShare's "Desktop Review"
15 April 2020
20 April 2020
21 April 2020
Who made the decision to purchase?
The Agreements
The First Agreement
Were the Purchase Order Terms and Conditions thereby incorporated into the First Agreement?
The Second Agreement
The article in The Guardian
Delivery of the Shangrila Ventilators
The technical tests - Dr Cook
The "Cancellation of Ordered Ventilators" brief
The clinical tests - Dr Gatward
The decision
The expert evidence
The contested functions of the Shangrila Ventilator
Pre-setting tidal volume
Volume control
The User Manual
Did the Shangrila Ventilators have "conventional" Volume Control?
The brochures
HealthShare's pleaded case
Dr Barkalow's evidence
Reliance
Pressure control mode
The User Manual
The MDT Shangrila Brochure
Accuracy of oxygen concentration
Expiry of oxygen sensors
Oxygen sensor calibration
Conclusion
Alarms
High Pressure Alarm
Circuit Disconnect Alarm
Other Alarms
Oxygen hoses
Patient circuit and expiratory valve calibration
Synchronised Intermittent Mandatory Ventilation
CPAP mode
Claim in contract - Sale of Goods Act
Did HealthShare accept the Shangrila Ventilators for the purposes of section 37 of the Sale of Goods Act?
Were the Shangrila Ventilators fit for purpose?
Particular purpose
Did MDT provide HealthShare with the MDT Shangrila Brochure and the User Manual merely as an intermediary?
Did HealthShare rely on MDT's skill or judgment - Ms Hanigan
Did HealthShare rely on MDT's skill or judgment - Dr Morris
Were the Shangrila Ventilators reasonably fit for the purpose of being used in a clinical setting for treating patients with COVID-19?
Was there a sale by description?
Misleading or deceptive conduct
Principles
The representations relied on
1. The ventilators were fit for the purpose of use by NSW Health in treating COVID-19 patients
2. The ventilators have a conventional volume-controlled ventilation mode and can reliably provide precise tidal volume
3. The ventilators would function in accordance with the specifications in the User Manual
4. The ventilators are intended to be used in common wards and ICUs
5. The ventilators are suitable for use in wards, emergency departments, ICUs or sub-acute care departments
6. The ventilators are capable of delivering FiO2 of between 40%-100% with ±20% accuracy
7. The ventilators can deliver volume control ventilation and the tidal volume on the ventilators can be pre-set
8. The ventilators trigger a continuous high-pressure alarm when the patient airway pressure exceeds a set limit
Conclusion